The NCDA&CS Solution Analysis Lab performs the following analyses based on standard industry methods:

  • Elements: Phosphorus (P), potassium (K), calcium (Ca), magnesium (Mg), sulfur (S), iron (Fe), manganese (Mn), zinc (Zn), copper (Cu), boron (B), sodium (Na), and aluminum (Al)
  • Nitrate-nitrogen (NO3-N)
  • Ammonium-nitrogen (NH4-N)
  • Chloride (Cl) 
  • pH
  • Soluble salts / Electrical conductivity 
  • Alkalinity 

Method detection limits (MDLs) are determined annually and when a new instrument or method is put into use. 

Minimum Sample Size

No matter how precise and accurate an analytical method is, meaningful results are only as good as the sample itself.  The most significant cause of poor statistical results in solution analysis is due to insufficient sample collection rather than analytical measurement.

To obtain a representative sample, NCDA&CS strongly recommends a sample volume of 8-16 oz (~500 mL).  Where this is not possible, please note the minimum sample volume required to perform each analysis.

AnalyteMiniumum Volume
Nitrate15 mL
Ammonium
Chloride
Elements (P, K, etc.)15 mL
pH50 mL
Electrical Conductivity
Alkalinity

Methods

Samples are analyzed as-received for pH, EC and alkalinity. For elemental and ion analysis, they are first filtered through a pre-folded Advantec #2 filter paper.  Prior to analysis, samples are homogenized by shaking. Except during analysis, samples are refrigerated at 4°C.

NO3-N is determined by cadmium reduction, where nitrate is reduced to nitrite with copperized cadmium, under alkaline conditions. The NO2-N concentration (that originally present plus reduced nitrate) is determined by diazotizing with sulfanilamide and coupling with N-(1-Naphthyl) ethylenediamine dihydrochloride to form a magenta-colored azo dye which is measured at 520 and 600 nm (USEPA 1993a; FIA NO3-W-1-1). 

NH4-N is determined by the modified Berthelot reaction where hypochlorite and sodium salicylate react with ammonia in a two-step reaction, converting it to 5-aminosalicylate.  The aminosalicylate is oxidized in the presence of sodium nitroferricyanide to form a blue-green colored complex, which is then measured at 660 nm (USEPA 1993b; FIA NO3-W-1-2). 

Both NO3-N and NH4-N are quantified by flow injection analysis (FIAlyzer-1000, FIA Lab; Seattle, WA). Nitrate-nitrogen (NO3-N) and nitrite-nitrogen (NO2-N) are reported as NO3-N and ammonium-nitrogen (NH3-N + NH4-N) is reported as NH4-N. 

Quality Controls. Samples are quantified using nine calibration standards.  A method blank (DI water, filtered) is analyzed with each batch.  A duplicate aliquot of a filtered solution sample is spiked and analyzed for analytical recovery with each batch. A calibration verification solution and calibration blank are analyzed at the beginning and end of each batch and after every 10 samples. Four independent calibration verification solutions are analyzed at the beginning and end of each run. Drift checks are analyzed at the beginning and end of each run and every 20 samples. Two nitrite checks (NO2-N) are analyzed to verify the completeness of the nitrate reduction reaction at the beginning and end of each run. 

Cl- is determined by the thiocyanate displacement method where thiocyanate is liberated from mercury(II) thiocyanate by the formation of soluble mercuric chloride. The liberated thiocyanate forms a red colored complex with ferric iron ions also present in solution (USEPA 1978a; FIA CL-W-1-1) This complex is quantified at 480 nm by flow injection analysis (FIAlyzer-1000, FIA Lab; Seattle, WA). Results are expressed in mg L-1.

 

Total elemental concentrations are determined by Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES) (Spectro Arcos EOP and Arcos II EOP, Spectro Analytical: A Division of Ametek; Mahwah, NJ) (USEPA 2001). Elements are measured at the wavelengths listed in Table 2 of NCDA&CS Solutions & Water Analysis Methods.

Quality Controls. A method blank (DI water, filtered) and calibration blank are analyzed with each batch. A calibration verification solution is run after the daily calibration, after every 10 samples, and at the end of each run. An independent calibration verification solution is analyzed at the beginning and end of each run. 

The pH of solution samples is determined directly on samples at room temperature using a Mettler Toledo T9 autotitrator and InMotion Max auto-sampler with a DGi115-SC pH electrode and DT1000 temperature probe (Mettler-Toledo, LLC; Columbus, OH) (APHA 2012; AOAC 1990c). 

Quality Controls. A three-buffer calibration is performed daily with a slope maintained between 98% and 102%. The pH 7 buffer is read back at the beginning and end of each day and every 10 samples.  A duplicate solution sample is analyzed daily as a quality control sample with an acceptance criterion of ±0.2 pH units. 

Electrical conductivity (EC) is a measure of the ability of an aqueous solution to carry a current. The EC of an aqueous solution depends on the total concentration, mobility, and valence of ions and on the temperature of the sample. EC is measured directly on samples at room temperature using a Mettler Toledo T9 autotitrator and InMotion Max autosampler with a InLab 731-ISM conductivity sensor and DT1000 temperature probe (Mettler-Toledo, LLC; Columbus, OH) (AOAC 1990b; USEPA 1982). EC is expressed in units of mS cm-1.  EC is also referred to as soluble salts (SS). 

Quality Controls. The calibration of the conductivity probe is checked daily with two EC checks.  A 500 µS/cm check is analyzed at the beginning and end of each day.  A 1413 µS/cm check is analyzed at the beginning and end of each day and every 10 samples. A duplicate solution sample is analyzed daily as a quality control sample with an acceptance criterion of < 2% relative standard deviation.

Alkalinity is a measure of water’s ability to resist change in pH (aka pH buffering capacity) and is the sum of carbonate (CO3) and bicarbonate (HCO3) ions in solution.  Carbonate (CO3) and bicarbonate (HCO3) concentrations are measured using a Mettler Toledo T9 autotitrator and InMotion Max autosampler with a DGi115-SC pH electrode and DT1000 temperature probe (Mettler-Toledo, LLC; Columbus, OH).  The autotitrator conducts titration over the carbonate/bicarbonate pH range using a weak acid and determines the inflection point(s) of the titration curve.  The amount of acid needed to lower the pH to the endpoint is inversely related to the alkalinity of the water. CO3 and HCO3 are reported in meq L-1. Total alkalinity (TA) is reported in ppm of calcium carbonate (CaCO3) and is calculated as follows: 

TA = (CO3 + HCO3) x 50

Quality Controls. A three-buffer calibration is performed daily with a slope maintained between 98% and 102%. The pH 7 buffer is read back at the beginning and end of each day and every 10 samples.  A 100 ppm CaCO3 titration check is measured at the beginning and end of each analysis.  A duplicate solution sample is analyzed daily as a quality control sample with an acceptance criterion of <3% relative standard deviation.

Association of Official Analytical Chemists. (1990a). AOAC official method 920.194: carbonate and bicarbonate in water—titrimetric method, final action. In: Official methods of analysis. 15th ed. Arlington (VA): AOAC International. P. 322.

American Public Health Association. (2012). Standard Methods for the Examination of Water and Wastewater. 22nd Edition. SM 4500 H+: pH Value. Ch. 4-91. Rice EW, Baird RB, Eaton AD, Clesceri LS, editors. Washington (DC): Public Health Association, Water Environment Federation, and American Water Works Association. 

FIA Lab.  CL-W-1-1. Method for Chloride Determination by Ferric Thiocyanate, Version 7.

FIA Lab.  NO3-W-1-1. Method for Nitrate Determination, Version 2.

FIA Lab. NH3-W-1-2. Method for Ammonia Determination by Salicylate Method, Version 1.

United States Environmental Protection Agency. (1993a).  Method 353.2, Revision 2: Methods for Determination of Nitrate-Nitrite Nitrogen by Automated Colorimetry. 

United States Environmental Protection Agency. (1993b). Method 350.1. Determination of Ammonia Nitrogen by Semi-Automated Colorimetry.

United States Environmental Protection Agency. (1978). Method 325.2: Methods for the Chemical Analysis of Water and Wastes. Chloride (Colorimetric, Automated Ferricyanide AAII). 

United States Environmental Protection Agency. (1982). Method 120.1. Methods for the Chemical Analysis of Water and Wastes. Conductance (specific conductance, μmhos/cm at 25 °C). 

United States Environmental Protection Agency. (2001). Method 200.7: Trace elements in water, solids, and biosolids by inductively coupled plasma–atomic spectrometry, revision 4.4. 